Basic Info.
Model NO.
B3
Transport Package
Wooden Case
Specification
2120*900*450-750mm
Trademark
YSENMED
Origin
China
Production Capacity
50, 000 Units/Year
Product Description
Baby Incubator Infant Care Equipment Comen B3 Infant Incubator Comen Neonatal Incubator B3 Baby Warmer Incubator
Neonatal incubator is warming equipment used for providing constant temperature and humidity for treatment of premature infants and critical neonates on the basis of satisfying their physiological characteristics and needs.
COMEN B3 incubator was thoughtfully designed to provides stable temperature and suitable humidity ambient, in addition to configure vital signs monitoring function to promoting the growth and stability of even the most fragile of patients.
Standard Configuration: Humidity Control, Temp (body temperature& incubator temperature), O2 Monitoring,
Optional Configuration: Weight Scale, Apnea Monitoring, Apnea awaking, Servo oxygen supply, Damping Doors
Safety Standards: ISO 13485:2003 approved, MDD 93/42/EEC as amended by MDD 2007/47/EC
Operation Environment
Working Temp 20~30°C
Humidity ≤80%
Air flow: ≤0.3 m/s
Power Supply 100-240V~, 50/60Hz±1Hz
Battery Type Rechargeable Lithium-ion battery
Battery Capacity 4400mAh
Battery Recharging Time: Maximum 6 hours for charging;
Battery backup 1 hours for continuous working (such functions as heating, humidifying shall be excluded)
Indicator: One alarm indicator, Power indicator, Battery indicator, QRS beep and alarm sound, Operating key sound
Interface: Parameter cable interface, AC power input, Three auxiliary power output, Plug-in slot, USB port
Date storage:
Alarm Review: 200 groups
Power-off storage: Yes
Alarm: User-adjustable High and Low 3-level Limits; Prioritized audible and visual alarm
Network: Connected to Central Monitoring System by hardwire/wireless
Temperature Control
Setting range: 35.0~37.5°C (Infant mode); 25.0~39.0°C (Air Mode), Set temperature exceeds 37°C, should press the display the "> 37°C" key, and then confirmed by the interface pop-up window to enter
Measurement range: 25.0~45.0°C
Resolution: 0.1°C
Accuracy: ±0.3°C
Alarm range: 34~38.5°C (Infant mode), 23~41°C (Air Mode)
Heating time: ≤35mins (Temperature inside the thermostat cover rises by 11°C)
Humidification time: ≤50mins (In indoor environment to 25°C±2°C, relative humidity of 40%±10% conditions, reach to 36°C, relative humidity of 85%)
Operation without adding water time: ≥12 hours
Humidity Control
Relative humidity range: 30%~95%
Resolution: 1%
Accuracy: ±10% (Humidity less than 85%); ≥75% (Humidity over than 85%)
Alarm: Audible and visual alarm; alarm events reviewable
Features
Air flow rate inside thermostat cover: ≤0.2m/s
Work mode: Air/ Infant
Bed tilt range: -10 ~ +10°
Input voltage 220 V~
Input current 6.6 A
Input frequency 50/60 Hz
Auxiliary output supply
Output voltage 240 V~
Output frequency 50/60 Hz
Output current 0.9A
Oxygen Servo-Controller
Setting range: 21~60%
Measurement range: 10~100%
Accuracy: ±2%
O2 supply pressure: 0.27~0.7MPa
O2 Concentration Calibration Available
Weight Scale
Measurement Range: 300g~8000g
Resolution: 1g
Accuracy: ±10g
Trend View: 7days
Neonatal incubator is warming equipment used for providing constant temperature and humidity for treatment of premature infants and critical neonates on the basis of satisfying their physiological characteristics and needs.
COMEN B3 incubator was thoughtfully designed to provides stable temperature and suitable humidity ambient, in addition to configure vital signs monitoring function to promoting the growth and stability of even the most fragile of patients.
Standard Configuration: Humidity Control, Temp (body temperature& incubator temperature), O2 Monitoring,
Optional Configuration: Weight Scale, Apnea Monitoring, Apnea awaking, Servo oxygen supply, Damping Doors
Safety Standards: ISO 13485:2003 approved, MDD 93/42/EEC as amended by MDD 2007/47/EC
Operation Environment
Working Temp 20~30°C
Humidity ≤80%
Air flow: ≤0.3 m/s
Power Supply 100-240V~, 50/60Hz±1Hz
Battery Type Rechargeable Lithium-ion battery
Battery Capacity 4400mAh
Battery Recharging Time: Maximum 6 hours for charging;
Battery backup 1 hours for continuous working (such functions as heating, humidifying shall be excluded)
Indicator: One alarm indicator, Power indicator, Battery indicator, QRS beep and alarm sound, Operating key sound
Interface: Parameter cable interface, AC power input, Three auxiliary power output, Plug-in slot, USB port
Date storage:
Alarm Review: 200 groups
Power-off storage: Yes
Alarm: User-adjustable High and Low 3-level Limits; Prioritized audible and visual alarm
Network: Connected to Central Monitoring System by hardwire/wireless
Temperature Control
Setting range: 35.0~37.5°C (Infant mode); 25.0~39.0°C (Air Mode), Set temperature exceeds 37°C, should press the display the "> 37°C" key, and then confirmed by the interface pop-up window to enter
Measurement range: 25.0~45.0°C
Resolution: 0.1°C
Accuracy: ±0.3°C
Alarm range: 34~38.5°C (Infant mode), 23~41°C (Air Mode)
Heating time: ≤35mins (Temperature inside the thermostat cover rises by 11°C)
Humidification time: ≤50mins (In indoor environment to 25°C±2°C, relative humidity of 40%±10% conditions, reach to 36°C, relative humidity of 85%)
Operation without adding water time: ≥12 hours
Humidity Control
Relative humidity range: 30%~95%
Resolution: 1%
Accuracy: ±10% (Humidity less than 85%); ≥75% (Humidity over than 85%)
Alarm: Audible and visual alarm; alarm events reviewable
Physical Characteristics | |
Size | 970mm×570mm×1350mm |
Screen Size | 8" TFT touch screen |
Weight | 100kg |
Resolution | 800 × 600 |
Water Tank Capacity | ≥1L |
X-ray tray | Available |
Humidifier | Available |
Drawer | Available |
Tray and bracket | Optional |
Caster wheel | four brakes |
Air flow rate inside thermostat cover: ≤0.2m/s
Work mode: Air/ Infant
Bed tilt range: -10 ~ +10°
Input voltage 220 V~
Input current 6.6 A
Input frequency 50/60 Hz
Auxiliary output supply
Output voltage 240 V~
Output frequency 50/60 Hz
Output current 0.9A
Oxygen Servo-Controller
Setting range: 21~60%
Measurement range: 10~100%
Accuracy: ±2%
O2 supply pressure: 0.27~0.7MPa
O2 Concentration Calibration Available
Weight Scale
Measurement Range: 300g~8000g
Resolution: 1g
Accuracy: ±10g
Trend View: 7days